Document Type
Article
Publication Date
2-2026
Keywords
Phantoms, Gels, Multi-layer phantoms, Tissue mimics, 3D printing, Resin printing, Vat photopolymerization
Abstract
Purpose Cardiovascular phantoms are used in biomedical research and development applications, allowing for complex geometries to be studied in a controlled environment. The various layers of human tissue have been difficult to mimic in these phantoms. In this study, a novel dual-layer cardiovascular phantom is created. Methods The interior lumen is 3D printed using an elastic vat-photopolymerization resin and cast within an industry standard tissue-mimicking ballistics gel. Strips of the 3D-printed resin were prepared and tested to determine Young's modulus, Ultimate tensile strength, and elongation at break. Results The final phantoms were reproducible, semi-transparent, and suitable for microCT scanning. Additionally, the 3D-printed elastic materials had: Young's Modulus of 12 +/- 3.2 MPa, UTS of 1.27 +/- 0.44 MPa, and elongation at break of 29 +/- 9%. These results are within the physiological ranges of human tissues. There was a moderate correlation between the thickness of the sample and stiffness, which may be important depending on the application of the models. Conclusion The methods for producing a dual-layered phantom are reproducible and appropriate for a variety of biomedical applications.
Citation
Elmer, K.M., Caffin, C., Scott, B. et al. Development and Characteristics of a Dual-Layered Vascular Phantom. Cardiovasc Eng Tech 17, 25–40 (2026). https://doi.org/10.1007/s13239-025-00810-0